ISO 16796:2022 pdf download – Nuclear energy — Determination of Gd 2 O 3 content in gadolinium fuel blends and gadolinium fuel pellets by atomic emission spectrometry using an inductively coupled plasma source (ICP-AES)

ISO 16796:2022 pdf download – Nuclear energy — Determination of Gd 2 O 3 content in gadolinium fuel blends and gadolinium fuel pellets by atomic emission spectrometry using an inductively coupled plasma source (ICP-AES).
To calibrate the measurement system for powder test samples, the standard solutions prepared in accordance with 8.1 are measured in at least duplicate and the calibration curve produced from the intensity-versus-concentration curve. To calibrate the measurement system for pellet test samples, the standard solutions prepared in accordance with 8.3 are measured in at least duplicate and the calibration curve produced from the intensity-versus-concentration curve. Calibration curves are evaluated using acceptance criteria established by each laboratory, based upon their historical performance; e.g., linear regression fit (R 2 ), slope, and coefficient of error. See example in Annex A. 9.3 Analysis of the test samples and quality control standards The test samples prepared in 8.2 and 8.3 shall be analysed after making the calibration curve. Test samples shall be measured in at least duplicate and spiked test samples should be measured at a defined frequency such as one test sample per batch of ten or fewer similar test samples. To avoid daily correction and achieve the required precision, a new calibration curve is prepared for each batch of test sample determinations. Process blank measurements should be performed periodically to ensure that background levels are in the expected range; however, background corrections are not normally necessary. Quality control (QC) standards shall be prepared and measured in the same manner as test samples. Compare the emission intensities of the test samples and QC standards to the calibration curve and calculate the Gd 2 O 3 concentrations for each of the test samples and QC standards. QC standards shall be used as both opening and closing controls, and should be control charted to monitor measurement and analyst performance. If any of the test samples contain products other than UO 2 and Gd 2 O 3 , the calibration and the instrument control measurements shall be performed using reference material blends having the same matrix as the test samples. 9.4 Measurement control program The measurement control program, procedure system, record retention program, method qualification, and staff training for measuring gadolinium by ICP-AES method shall conform with the accepted practices at the user laboratory. Guidance on record retention, method qualification, and staff training is available in ISO/IEC 17025 [2] . Guidance on a measurement control program is available in ANSI N15.51 [4] QC standards shall be traceable and of sufficient quality to demonstrate control of measurement parameters and performance/proficiency of analysts. Periodic spiking of a predetermined percentage of test samples with known quantities of gadolinium is recommended. Spike recoveries should be consistent with measurement uncertainty determined from quality control standards and should support the conclusion that the measurement method may be applied to demonstrate that materials meet product specification. Measurement of process blanks is recommended to detect unexpected gadolinium background levels.

Tags: Download